myPCProbe
NOVUS PRODUTOS ELETRONICOS LTDA www.novusautomation.com 2/11
Figure 02 – TEMP model dimensions
Ambient temperature channel: Internal thermistor.
Built-in cold junction compensation for thermocouples and wire resistance for Pt100.
Total accuracy:
Thermocouple K: ± 4.5°C (when maximum resoluti on is selected);
Pt100: ± 1.5°C (when maximum resolution is sel ected);
Ambient Temperature Channel: ± 1.5°C (20 minut es after connection to the USB port)
Note: It is possible to reduce (and even cut out) errors if after calibration the calibration points are
transferred to the device using the custom calibration screen (via software).
A/D resolution: Configurable from 15 to 11 bits.
Sampling rate: 8 to 128 samples per second, depending on resolution and number of channels enabled.
RHT Model (temperature and relative humidity)
Maximum measurement range: -20 to 100ºC and 0 to 100% RH (available also as dew point). See Figure 04
“Humidity sensor operating conditions”.
Dimensions: See Figure 03.
Figure 03 - RHT model dimensions
Accuracy: See Figure 04.
Note: It is possible to reduce (and even cut out) errors if after calibration the calibration points are
transferred to the device using the custom calibration screen (via software).